Procedure
1. Record the hand dipped ullage values and the corresponding values measured by
the level gauge.
2. Enter the hand dipped level values and the level gauge values into the WinSetup
Calibration Data
window (see
).
3. Inspect the resulting calibration graph and, if necessary, exclude measurement
points which should not be used in the adjustment calculation.
4.7.3
To enter calibration data
Prerequisites
Make sure that the following information is available when you intend to use the
Calibrate
function in Rosemount TankMaster WinSetup:
•
A list of hand dipped ullage values.
•
A list of level values measured by the Rosemount 5900S that correspond to the hand
dipped ullage/level values.
Procedure
1. In the Rosemount TankMaster WinSetup workspace window select the Rosemount
5900S level gauge to be calibrated.
2. Click the right mouse button and choose
Calibrate
, or choose
Calibrate
from the
Service/Devices
menu.
3. The
Calibrate
window is empty before any data is entered. Ensure that the gauge
communicates properly with TankMaster by verifying that the Tank Reference
Height appears in the lower left corner.
4. Click the
Calibration Data
button.
Configuration
Reference Manual
September 2019
00809-0100-5900
130
Reference Manual
Summary of Contents for Rosemount 5900S Series
Page 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Page 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Page 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Page 266: ...Product Certifications Reference Manual September 2019 00809 0100 5900 266 Reference Manual ...