3.3.2
Parabolic antenna
Mounting the clamped Flange Ball
Follow this instruction when installing the clamped Flange Ball on a flange.
Prerequisites
1. Use a flange of thickness 6 - 30 mm.
2. Make sure that the diameter of the hole is 96 mm. Make a small recess at one side
of the flange hole.
Figure 3-17: Flange Requirements
A. Recess
Procedure
1. Put the O-ring on the flange and insert the Flange Ball into the hole. Make sure that
the guide pin on the side of the Flange Ball fits into the recess on the flange.
$
%
A. Flange Ball
B. Nut
Installation