8. If the position deviates from the Nominal Position, return to the
LPG Configure Pins
window and enter the measured position in the
Nominal Position
field.
Note
The nominal position that was entered the first time refers to the mechanical
distance. The measured position refers to the electrical distance which is the
distance “seen” by the level gauge.
9. Repeat
#unique_145/unique_145_Connect_42_Step7
Successful Verification
appears, indicating that Nominal Position
corresponds to Measured Position.
Reference Manual
Configuration
00809-0100-5900
September 2019
Rosemount 5900S Radar Level Gauge
127