Table 3-9: Conductor Cross-sectional Area
Conductor connection
Cross-section
Minimum
Maximum
Solid
0.2 mm
2
/ AWG 24
1.5 mm
2
/AWG16
Flexible
0.2 mm
2
/ AWG 24
1.5 mm
2
/AWG16
With wire end ferrule
0.25 mm
2
/ AWG 24
1.5 mm
2
/AWG16
With plastic collar ferrule
0.25 mm
2
/ AWG 24
0.75 mm
2
/AWG19
In case conductor insulation diameter exceeds 2.9 mm it may not be possible to insert
cable properly into the terminal block. In such a case the stripping length may have to be
increased. Adjust stripping length such that no bare conductor appears outside the
terminal when conductor is attached to the terminal block.
A solid conductor, or a flexible conductor with end ferrule, can easily be pushed into the
terminal block without using any tools. In case a flexible (stranded) conductor is used, you
will have to push the release button in order to insert the conductor.
Figure 3-25: Push the Conductor Into the Terminal Block
$
%
&
A. Terminal block
B. Release buttons
C. Conductor
To disconnect, push the release button and remove the conductor.
Reference Manual
Installation
00809-0100-5900
September 2019
Rosemount 5900S Radar Level Gauge
85
Summary of Contents for Rosemount 5900S Series
Page 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Page 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Page 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Page 266: ...Product Certifications Reference Manual September 2019 00809 0100 5900 266 Reference Manual ...