3-103
Laser Address Selection
Caution!
To prevent damage to the laser heads, make sure the antennas do not come into
contact with obstacles (e.g., the ground or base plate) as they are being rotated.
Also, when performing an uncoupled alignment and the laser heads are
mounted in close proximity to each other, make sure the antennas do not come
into contact with each other as one laser head is rotated past the other.
Acquiring Auto 4 Point Data
If the laser heads can only be rotated to three of the cardinal positions, the reading
for the fourth will automatically be calculated when “Enter” is pressed after the
third reading has been acquired. Using only three readings increases the likeli-
hood of error and does not allow the instrument to check data validity. Emerson
does not recommend using only three readings if four are available.
Clear key
Use the Clear key to clear and restart the current averaged laser reading for each
laser head at any time.