3-47
Laser Address Selection
Dimensions can be expressed in either inches or millimeters (mm), depending on
the analyzer units. The range of values that can be entered for the A, C, and E
dimensions are 0.01 to 3600 inches and 0.254 to 91440 mm. The range of values
that can be entered for the B, D, and F dimensions are -3600 to 3600 inches and
--91440 to 91440 mm.
When entering any of the machine dimensions, they should be measured and
extended to the nearest 1/8 inch (3 mm) with the exception of dimension C.
Dimension C should be measured to the nearest 1/16 inch (1.5 mm). Refer to the
Entering Fractions section on page 3-51 for information about entering fractions.
In addition to the RPM, the C and F dimensions (if using standard tolerances) are
required to calculate and view the machine’s alignment condition available in the
Move Machine function. If any of these parameters have not been entered, a
warning message will be displayed before continuing. The warning message will
prompt you to verify that you want to proceed without the required information. If
you answer yes, then you will advance to the next step in the procedure. Also, if the
C dimension is missing and the Foot Pre-Check option has been enabled, it will be
disabled before advancing. If you answer no, then the operation is aborted and you
remain at the Enter Dimensions screen.
If only the RPM and C and F dimensions have been entered, then the Quick Spec
mode will be activated. In Quick Spec mode only the Angle/Offset data will be
available, no machine moves will be calculated, and if either the thermal growth or
Review Results options have been enabled they will be disabled. In this case, a
warning message displays before continuing. The warning message prompts to
verify that proceeding to Quick Spec mode is desired. If you answer yes, then it
advances to the next step in the procedure. If you answer no, then the operation
aborts and the unit remains at the Enter Dimensions screen.
In addition to the RPM and C and F dimensions, if any of the following dimensions
are missing the Quick Spec mode activates. Refer to the Quick Spec section on
page 3-53 for more information.
The A, B, C, D, and E dimensions are required to calculate and view the six alter-
nate machine move solutions that are available in the Move Machine function.
• A, B and D
m
•
B, D and E
•
B and D
•
A, B and E
•
A and D
•
B and E
•
A, D and E
•
A and E