4. REMOTE OPERATION
Page 165
© 1998-2008 DH Instruments, a Fluke Company
VAC(=)
Purpose
To set or read PG7000 reference mode. This command is only valid
for PG7601.
Syntax “VAC=n”
“VAC”
Default VAC=0
Argument
n = 1:
n = 0:
PG7000 is being operated with a bell jar and vacuum
PG7000 is being operated at atmospheric pressure
Remarks
You must let PG7000 know when you are operating under vacuum. The reported
PG7000 measured pressure will then reference the vacuum.
Example Typical
command:
Typical reply:
“VAC=1”
“VAC=1”
Error
See 9.3.2 Error Messages
ERR #1
If n not set to 1 or 0
See Also
3.9.4, MODE
VENT(=)
Purpose
To vent the test pressure to atmosphere if an automated pressure generation/control
component is being used for automatic pressure generation.
Syntax “VENT=n”
Default VENT=0
Argument
n = 1:
n = 0:
Activates the vent procedure on an automated pressure
generation/control component
Closes exhaust valve on an automated pressure
generation/control component
Remarks
When n = 1 the external automated pressure generation/control component test
pressure will quickly decrease to atmospheric.
Example Typical
command:
Typical reply:
“VENT=1”
“VENT=1”
Error
See 9.3.2 Error Messages
ERR #1
If n not set to 1 or 0
See Also
3.9.9, PPC, PGEN
VER
Purpose
Read the version number of the internal software.
Syntax “VER”
Default N/A
Argument N/A
Remarks
The software version of the EPROM can be read.
Example Typical
command:
Typical reply:
“VER”
“DH INSTRUMENTS, INC PG7302 Ver2.00”
Error None
VOL
Purpose
Read or set the currently active test volume for AutoGen when using a PG7302 or
PG7307 with a PPCH pressure controller
Syntax “VOL=
cc
”
“VOL”
Default “VOL=30”
Argument
cc
: The test volume in cubic centimeters (0 to 300)
Remarks
This setting only affects operation of a PG7302 or PG7302 (oil operated piston gauge)
when using automated pressure control (AutoGen) with a PPCH pressure controller.
Accurately specifying the test volume improves pressure control for AutoGeneration of
initial piston float and for refloating. It is indispensable if the test volume is greater than
100 cc.
Example Typical
command:
Typical reply:
“VOL=50”
“50”
Error
ERR #1 Invalid argument
See Also
3.9.9, 3.9.9.6