3. GENERAL OPERATION
Page 127
© 1998-2008 DH Instruments, a Fluke Company
3.11.8.3
<2CONTROL>, <3LOADALL>
PURPOSE
To clear the AMH mass handler and cause all the masses in the AMH mass set to
be loaded onto the piston. The function is used to remove all the masses from the
AMH mass handler and prepare it to be removed or installed on the PG7000
platform.
OPERATION
To load all AMH masses, press
[SPECIAL]
,
<8AMH>
,
<3loadall>
. Confirm the
loadall instruction and AMH proceeds to lower all the masses onto the piston.
Do not put your fingers or anything else under the AMH trim mass tray. When
the mass load is lowered, it will pinch or crush anything under it.
3.11.8.4
<2CONTROL>, <4UNLOADALL>
PURPOSE
To clear the AMH mass handler and cause all the masses in the AMH mass set
to be unloaded from the piston (retained in the AMH automated mass hander).
OPERATION
To unload all AMH masses, press [SPECIAL], <8AMH>, <4unloadall>. Confirm the
unloadall instruction and AMH proceeds to raise all the masses off the piston. When
the mass handler lowers, the bell and mass lifting shaft are still loaded on the piston.
Do not put your fingers or anything else under the AMH trim mass tray. When
the mass load is lowered, it will pinch or crush anything under it.
3.11.9
<9RESET>
PURPOSE
To reset various PG7000 settings to default or factory values.
PG7000 stores its user definable settings in non-volatile memory. The reset menu allows the
user to selectively or completely reset these settings to factory defaults. Resets clear
settings that the user may have made, and should be used only to restore the PG7000 to a
known state. PG7000 will go through its reboot routine after any type of reset is executed.
OPERATION
To access the reset choices press
[SPECIAL]
and
select
<9reset>
. The display is:
1sets 2units 3com
4cal 5setups 6all
Select the desired reset. After confirmation, the reset occurs. A reset always puts PG7000
through its start up routine as if power had been turned OFF and back ON.
See Sections 3.11.9.1 through 3.11.9.6 for detailed information on the specific reset choices.
Reset functions change user settings that affect pressure measurement. If not used
properly, resetting can cause out of tolerance measurements. Reset functions should only be
used by qualified personnel with reference to this manual for information on the reset functions.