3. GENERAL OPERATION
Page 51
© 1998-2008 DH Instruments, a Fluke Company
OPERATION
To customize the
[UNIT]
function key, from the main run screen press
[SETUP]
and select
<2PresU>
. The display is:
1.
The UNIT number in the
[UNIT]
menu that is
to be changed.
Set up user unit #6
Enter the number of the unit position that
you would like to change. The display
becomes:
Unit #6 1SI 2other
3altitude 4user
Select the desired pressure unit category. SI units include units
based
on SI
such as mmHg. Select the desired unit from the unit list (see Table 15).
Table 15
.
Pressure Units of Measure Available
<1SI>
<2Other>
<3altitude>*
<4User>**
<1Pa>
<2Kpa>
<3MPa>
<4mbar>
<5bar>
<6mmHg>
<7mmWa>
<1psi>
<2psf>
<3inHg>
<4inWa>
<5kcm2>
<6Torr>
<7mTor>
<8none>
<1m>
<2ft>
<1user>
*
<3altitude>:
PG7601 in absolute by vacuum mode only. Not available in
PG7102, PG7202 and PG7302.
**
<4user>:
User defined unit. As there are no altitude units on PG7102,
PG7202 and PG7302, user is <
3user>
.
The UNIT function display is not required to display six units. To delete the
current unit from the UNIT screen and show no unit, select
<2other>
,
<6none>
for that unit number.
If
<4user>
was selected, the user unit must be defined. The display is:
1. Entry
field.
Define user unit:
1.000000 Units/Pa
Enter the number of user units per Pascal [Pa] in the entry field. Pressing
[ENTER]
defines the user unit and returns to the
<Set up unit #>
screen.
See Section 7.1.1 for the pressure unit of measure conversion factors used
by PG7000.
The user defined unit can be assigned a user defined label using the UDU
remote command (see Section 4.3.4.2).