3. GENERAL OPERATION
Page 63
© 1998-2008 DH Instruments, a Fluke Company
1. Tare PG7000 (PG7102, PG7202, PG7302, or
PG7601)
2. Differential Device Under Test (DUT
3. Reference PG7000 (PG7102, PG7202 or
PG7302)
4. MPC1-D or other INLET Valve
5. MPC1-D or other Hi Variable Volume (Hi VV)
6. MPC1-D or other VENT Valve
7. MPC1-D or other HI/LO BYPASS Valve
8. MPC1-D or other Lo Variable Volume (Lo VV)
Figure 11
.
High Line Differential Mode Schematic
DHI Technical Note 0080TN03 provides more detailed information on high line
differential mode principles using PG7102 and PG7601 piston gauges and includes a
complete uncertainty analysis. Consult DHI or visit our website,
www.dhinstruments.com to obtain a copy.
OPERATION
High line differential mode does not support operation with AMH automated
mass handler(s).
To select high line differential mode operation and access differential mode
functions press
[MODE]
,
<3HLdif>
on the
reference
PG7000.
When operating in high line differential pressure mode, the reference PG7000
is “master” and the tare PG7000is “slave”. All operator interaction is with the PG
Terminal of the reference PG7000. The tare’s display is for information only. The
tare PG7000 keypad is inactive.