PG7000™ OPERATION AND MAINTENANCE MANUAL
© 1998-2008 DH Instruments, a Fluke Company
Page 118
OPERATION
PG7000 is delivered with NO active password so access to the User Level
menu is open. The User Level is set to <1Low>. User Levels can be changed freely
until a password has been created. Reset functions (see Section 3.11.9) do not
affect the password setting.
To access the User Level function, press
[SPECIAL]
and select
<2level>
.
If NO password yet exists or if the
correct password has been entered
, the
display is:
1change user level
2edit password
Selecting
<1change user level>
brings
up the restriction menu:
Restriction: 1none
2low 3medium 4high
You can then select the current restriction level, or press
[ESCAPE]
to return to
the main run screen.
Selecting
<2edit password>
displays the
user password and allows it to be edited.
Passwords can be up to six numbers in
length and cannot start with a zero.
Password:
pppppp
0 disables password
If
0
is entered as the password value, then the password is made inactive and a
password will NOT be required to access the User Level menu. This is the
factory default with a security level of
<2low>
.
Once a password has been entered, the User Level cannot be changed without
reentering the password.
If there is an active password
, the
PG7000 password entry screen appears.
PG7000 NS
nnn-xx
Password:
pppppp
The user must enter the user defined password or the factory
secondary
password to proceed. When a password is entered correctly, operation
proceeds to the
<1change user level 2edit password>
screen.
The first field,
<nnnn>,
is the serial number of PG7000, followed by a second
field,
<xx>.
That counts the number of times that a
secondary
password has
been used. This second field increments each time a
secondary
password
is used. The third field,
<pppppp>
, is for normal password entry.
The factory
secondary
password is available in case the user’s password has
been misplaced or forgotten. It can be obtained by contacting a
DHI
Authorized
Service Provider. The factory
secondary
password is different for all PG7000’s
and changes each time it is used.
3.11.5
<5REMOTE>
PURPOSE
To configure the PG7000 COM1, COM2, COM3 and IEEE-488 communication ports. To test
COM1, COM2 and COM3 communications.