3. GENERAL OPERATION
Page 109
© 1998-2008 DH Instruments, a Fluke Company
PRINCIPLE
See Section 3.11.1.6, PRINCIPLE.
OPERATION
Mass loading bell information is element specific metrological data.
Uninformed or accidental altering of this information may lead to out of tolerance
measurements. Mass loading bell information should only be edited by qualified
personnel as part of the PG7000 calibration or recalibration process.
PG7000 can store up to 3 mass bell files. When <3add> is selected and all
the files have already been used, a warning is displayed. A mass loading bell must be
deleted before a mass loading bell can be added.
To create a new PG7000 mass loading bell press
[SPECIAL]
and select
<1PC/MS>
,
<3mass bell>
,
<3add>
.
Edit the serial number to the number of
the mass bell being added and
press
[ENTER]
. The display is:
Add mass bell:
S/N 1
Edit the average mass density of the
mass loading bell and press
[ENTER]
.
The value defaults to the default density
for the typical mass loading bell for the
PG7000 model. The display is:
Mass density?
5058 kg/m3
The AMH automated mass handler mass bell includes the bell, the binary
mass carrier and the mass lifting shaft.
Edit the calibration report number (up to
nine digits) to the the number of the
current calibration report of the mass
loading bell that is being added and press
[ENTER]
. The display is:
Cal report number?
1775
Edit the calibration report date to the date
of the current calibration report of the
mass loading bell that is being added and
press
[ENTER]
. The display is:
Cal report date?
20040101
Edit the the true mass value of the mass
loading bell that is being added and press
[ENTER]
. The value defaults to the nominal
mass of the typical mass loading bell for the
PG7000 model. The display is:
Bell #2832
True mass: 0.449985
For high line differential mode, when setting up the tare bell of the tare
PG7000, the true mass of is entered as equal to the nominal mass. Enter the
nominal mass of the bell as the true mass.