PG7000™ OPERATION AND MAINTENANCE MANUAL
© 1998-2008 DH Instruments, a Fluke Company
Page 126
Select
<2control>
to control AMH directly:
Select
<1up/down>
to move the AMH mass handler up or down (see Section
3.11.8.1).
Select
<2discreet>
to specify masses to be loaded and unloaded (see Section
3.11.8.2).
Select
<3loadall>
to load all the AMH masses onto the piston (see Section 3.11.8.3).
Select
<4unloadall>
to unload all the AMH masses from the piston (see Section
3.11.8.3).
3.11.8.1
<2CONTROL>, <1UP/DOWN>
PURPOSE
To move the AMH mass load to the up or down position (see the AMH-38/
AMH-100 Operation and Maintenance Manual).
OPERATION
To move the AMH mass handler to the up and down positions, press
[SPECIAL]
,
<8AMH>
,
<1up/down>
.
The display indicates the current position. Press
[ENTER]
to change the
position.
Do not put your fingers or anything else under the AMH trim mass tray. When
the mass load is lowered, it will pinch or crush anything under it.
3.11.8.2
<2CONTROL>, <2DISCREET>
PURPOSE
To load and unload specifc AMH masses.
OPERATION
To load and unload specific AMH masses, press
[SPECIAL]
,
<8AMH>
,
<2control>
,
<2discreet>
.
The display is (64 is present only with
AMH-100):
1* 2 4* 8 16 32
64 Main:1
The
<1>
,
<2>
,
<4>
,
<8>
,
<16>
,
<32>
and
<64>
indicate the 0.1, 0.2, 0.4, 0.8, 1.6,
3.2 and 6.4 kg masses.
<*>
indicates that the mass is loaded. Move the cursor
to a mass and press
[+/-]
or
[.]
to change its status.
<Main:>
indicates the number of main masses that are loaded. Edit this number
to the desired value.
If any condition in the discreet display is changed the bottom right of the display
changes to
<ENT loads>
. Press
[ENTER]
to causes AMH to operate to load the
new mass configuration. Press
[ESCAPE]
to return to the AMH menu with no
change in mass load.
Do not put your fingers or anything else under the AMH trim mass tray. When
the mass load is lowered, it will pinch or crush anything under it.