PG7000™ OPERATION AND MAINTENANCE MANUAL
© 1998-2008 DH Instruments, a Fluke Company
Page 192
Figure 20.
Filling Gas Operated, Liquid Lubricated
Piston-Cylinder Module Reservoir (PC-7200)
When refilling piston-cylinder modules, if the quantity of liquid withdrawn in step above is
less than 2 cc, you should fill the module more frequently.
In day to day use, always keep gas operated, liquid lubricated modules in the vertical
position when handling them to avoid having liquid escape from the fill holes.
5.3.4
CLEANING PISTON-CYLINDERS
Water/Detergent Method
Of the two cleaning methods, the water/detergent method is more time consuming but it is
also more thorough.
DO NOT undertake piston-cylinder module disassembly without familiarizing yourself
with Section 5.3.1 of this manual. Incorrect disassembly may damage or destroy the piston-
cylinder element.
Disassemble the piston-cylinder module following the instructions given in Section
5.3.2.1 for PC-7100/7600 modules, Section or Section 5.3.2.2 for PC-7200 modules or
Section 5.3.2.3 for PC-7300 modules.
Reinsert the piston (with cap installed) into the cylinder.
Create a bath of water and mild liquid dishwashing detergent. Distilled water is
acceptable. De-ionized water is best. Detergent must be free of additives (i.e., lotions
or softening agents which may leave a residue after cleaning). To avoid undesired
thermal effects, bring the bath to room temperature.
Holding the piston-cylinder assembly by the outside of the cylinder and the piston cap,
submerge the assembly in the bath. Using a rotating motion, thoroughly work the
detergent into the piston-cylinder assembly.