PG7000™ OPERATION AND MAINTENANCE MANUAL
© 1998-2008 DH Instruments, a Fluke Company
Page 108
OPERATION
Once a mass set file has been deleted it cannot be recovered. To recreate it,
the add mass set function must be used (see Section 3.11.1.6) which requires
entering all the mass set information. Before deleting a mass set, consider editing
it (see Section 3.11.1.7).
To delete a mass set press
[SPECIAL]
,
<1PC/MS>
and select
<2mass set>
,
<4delete>.
Select the mass set to be deleted and press
[ENTER]
. Select
<2yes>
to delete the mass set. Press
<1no>
to NOT delete.
The active mass set (mass set that is currently selected) cannot be deleted.
Before attempting to delete a mass bell, be sure it is not currently selected.
3.11.1.10
SELECT MASS SET
PURPOSE
To select a mass set to be active and used by PG7000 in its mass loading and
defined pressure calculations. To initialize the optional AMH automated mass
handler if present.
OPERATION
To select the active mass set press
[SPECIAL]
,
<1PC/MS>
and select
<2mass
set>
,
<5select>
.
Mass sets for an AHM automated mass handler are identified by a letter
<A>
after their
serial number.
Put the cursor on the desired mass set and press
[ENTER]
.
If a MANUAL mass set is selected, operation returns to the MAIN run screen.
If an AMH automated mass handler mass set is selected, the PG7000 platform
attempts to establish communication with the AMH and initialize it. If the
communication cannot be established with the AMH, an error is displayed. If
communication with the AMH mass set is established, the AMH is initialized and all
masses are loaded. If the AMH is unable to complete initialization, an error message is
displayed. The most common reason for failure to complete initialization is inadequate
drive pressure supply to the AMH. For trouble shooting, see the AMH-38/AMH-100
Operation and Maintenance Manual. If the AMH initialization completes successfully,
operation returns to the main run screen with the AMH active. The AMH remains active
until a non-AMH mass set is selected. The PG7000 platform will attempt to initialize the
AMH on power up if an AMH mass set is the active mass set.
3.11.1.11
ADD MASS LOADING BELL
PURPOSE
To create a new mass loading bell that will be available for selection as the active
mass loading bell. Use this function when a new mass loading bell has been acquired.
If the mass loading bell and PG7000 Platform were delivered together, the
mass loading bell has already been entered at the factory. Before creating a new
mass loading bell, check whether it already exists (see Section 3.11.1.13).