PG7000™ OPERATION AND MAINTENANCE MANUAL
© 1998-2008 DH Instruments, a Fluke Company
Page 96
<1internal>:
Use real time readings from PG7000’s on-board piston-cylinder temperature
platinum resistance thermometer (embedded in the piston-cylinder module
mounting post).
<2normal>:
Use a fixed value of 20
°
C.
<3user>:
Use a fixed user entered value. If
<3user>
is selected, the user value must be
entered, in degrees Centigrade
[
°
C]
.
Select
<5g>
to specify the value of acceleration due to gravity that is used in calculating the
force applied to the piston. Pressing
<5g>
offers the gravity variable source choices:
<1locall>:
Use the fixed value of local gravity stored under
[SPECIAL]
,
<6gl>
(see Section
3.11.6).
<2normal>:
Use a fixed value of 9.806650 m/s
2
.
<3user>:
Use a fixed user entered value different from the current local gravity value
stored in the
[SPECIAL]
,
<6gl>
menu option. If
<3user>
is selected, the user
value must be entered, in meters per second squared [m/s
2
].
Select
<6READY>
to edit the values of
Ready/Not Ready
status criteria (see Section 3.4).
These include:
<1position>:
Piston position limits. Edit the fixed value, in millimeters
[mm]
, of the band
around mid-float position within which the piston position is
Ready
(see
Section 3.4.1). This also determines the limit at which the piston is refloated by
the GEN function when GEN is ON (see Section 3.9.9).
<2vacl>
(PG7601 only):
Maximum reference vacuum pressure when operating in absolute by vacuum
mode. Edit the value, in Pascal
[Pa]
, under which the reference vacuum must
be for a vacuum
Ready
condition to occur (see Section 3.4.3).
Select
<7vac>
to specify the reference vacuum variable source for the value used for the
pressure under the bell jar when calculating the absolute pressure defined in
absolute by
vacuum
mode. Pressing
<7vac>
offers the reference vacuum variable source choices:
<1internal>:
Use real time readings from PG7000’s on-board vacuum gauge (mounted
directly in the vacuum plate, under the mass load).
<2normal>:
Use a fixed value of 0 Pa.
<3user>:
Use a fixed user entered value. If
<3user>
is selected, the user value must be
entered, in Pascal
[Pa]
.
<4COM2>:
Use real time readings from or through the device connected by RS232 to
PG7601’s COM2 port. The external vacuum gauge must be properly set up to
communicate with PG7601 (see Section 3.11.5.5 for information on setting up to
read an external vacuum gauge).
3.11
[SPECIAL] MENU
PURPOSE
The
[SPECIAL]
key accesses a menu of PG7000 functions and settings that are less
commonly or not normally used in regular operation.
OPERATION
To access the SPECIAL menu, press
[SPECIAL]
from
the main run screen. This display is:
1PC/MS 2presU 3head
4prefs 5remote 6gl
7cal 8AMH 9reset
Some screens, such as the SPECIAL menu, go beyond the two lines provided by the display.
This is indicated by a flashing arrow in the second line of the display. Press the [
←
] and [
→
] keys
to move the cursor to access the lines that are NOT visible or directly enter the number of the
hidden menu choice if you know it.