PG7000™ OPERATION AND MAINTENANCE MANUAL
© 1998-2008 DH Instruments, a Fluke Company
Page 102
When the desired piston-cylinder module is identified, press
[ENTER]
to proceed
with viewing piston-cylinder module information. Successive pressing of
[ENTER]
steps through the piston-cylinder information screens in the same order
as the add and edit functions (see Section 3.11.1, View Piston-Cylinder
Module, Principle).
3.11.1.4
DELETE PISTON-CYLINDER MODULE
PURPOSE
To delete a piston-cylinder module file so that it is no longer included in the
piston-cylinder modules available for selection by pressing
[P-C]
.
OPERATION
Once a piston-cylinder module file has been deleted it cannot be recovered.
To recreate it, the add piston-cylinder module function must be used (see Section
3.11.1.1) which requires entering all the piston-cylinder module information. Before
deleting a piston-cylinder module, consider editing it (see Section 3.11.1.2).
To delete a piston-cylinder module press
[SPECIAL]
,
<1PC/MS>
and select
<1PC>
,
<4delete>.
The display identifies the currently active piston-cylinder module. Press
[ENTER]
to delete the currently active piston-cylinder module or press
[P-C]
to
toggle through the other piston-cylinder modules available (see Section 3.9.2).
When the desired piston-cylinder module is identified, press
[ENTER]
to proceed
with deleting the piston-cylinder module information. Confirmation to delete is
required. Select
<2yes>
to delete the piston-cylinder module. Select
<1no>
to
NOT delete.
The active piston-cylinder module (piston-cylinder module that is currently
selected) cannot be deleted. Before attempting to delete a piston-cylinder module,
be sure it is not currently selected.
3.11.1.5
SELECT THE ACTIVE PISTON-CYLINDER MODULE
PURPOSE
To select the piston-cylinder module that is active and used by PG7000 in its
mass loading and defined pressure calculations.
This function serves the same purpose as pressing [P-C] (see Section 3.9.2).
OPERATION
To select the active piston-cylinder module press
[SPECIAL]
,
<1PC/MS>
and
select
<1PC>
,
<5select>
.
3.11.1.6
ADD MASS SET
PURPOSE
To create a new mass set that will be available for selection as the active mass
set to load on the PG7000 piston. Use this function when a new mass set has
been acquired or a mass set was deleted and must be reentered.