4. REMOTE OPERATION
Page 151
© 1998-2008 DH Instruments, a Fluke Company
Error
See 9.3.2 Error Messages
ERR
#1
ERR #2
ERR #3
ERR #29
ERR #30
x mass set # is invalid
The nominal mass value given is invalid
The true mass value not within 10% of the nominal mass
The mass set has not been “opened” yet
You are at the end of the mass set
See Also
3.11.1.6, 3.6, 2.3.1.3, MASS, MASSx
Error
ERR #1 Invalid mass load or AMH not active.
MEDIA(=)
Purpose
Read or set the PG7000 pressurized medium type for DUT head corrections.
Syntax “MEDIA=mediaType”
“MEDIA”
Default “MEDIA=N2
Argument
mediaType:
“N2”, “He”, “Air”, “Oil”, “H2O” or user set density value.
Remarks
You must select one of five (5) available media for use with the PG, of specified a user
defined density. This allows PG7000 to utilize the correct internal calculations for head
corrections.
Example Typical
command:
Typical reply:
“MEDIA=Air”
“Air”
Error
See 9.3.2 Error Messages
ERR #1
Media type is invalid
See Also
3.9.7 PRINCIPLE, 3.11.3.1
MEM
Purpose
Read the status of the internal data RAM since the last power up.
Syntax “MEM”
Default N/A
Argument N/A
Remarks
On power up a memory test is run to check the integrity of the internal data RAM. If the
memory has been corrupted then “FATAL MEMORY LOSS” will be displayed to alert the
user. The status memory can be read from a remote computer.
Typical causes of memory fault:
Upgrade of PG7000 software
Connection of a cable to PG7000 when PG7000 is on
Failure of internal memory
Return
string:
“MEM=1”
System is OK
“MEM=0”
System memory has been corrupted and the default operating
parameters were loaded into memory
Example Typical
command:
Typical reply:
“”MEM”
“MEM=1”
Error None
MMODE
Purpose
Get or set the PG measurement mode and reference.
Syntax “MMODE”
“MMODE=
mode
”
Argument
Mode:
"AATM" Absolute mode with atmospheric reference
"AVAC Absolute mode with vacuum reference
"G" Gauge mode with atmospheric reference
"D" Differential mode (PG7601) or high line differential mode
(PG7102)
Remarks
This command sets both the unit mode and the measurement reference. This can also
be accomplished using the UNIT and the VAC commands. Differential mode is a
special case depending on the PG7000 model, and requires special conditions to run.
Example Typical
command:
Typical reply:
“MMODE”
“AATM”
Error ERR
#1
ERR #32
Invalid mode specified
Altitude or airspeed unit not allowed with atmospheric reference.
See Also
3.9.4, 3.9.3, UNIT, VAC