3. GENERAL OPERATION
Page 93
© 1998-2008 DH Instruments, a Fluke Company
OPERATION
To access the SETUP menu, press
[SETUP]
from the main run
screen. The display is:
Current SETUP: #01
1select 2view 3edit
#nn in the upper right hand corner displays the number of the SETUP file that is currently active.
Select
<1select>
to select a different active SETUP file. Select
<2view>
to view the variable choices and
values of the active SETUP file. Select
<3edit>
to create or edit a SETUP file.
See Sections 3.10.1 to 3.10.3 for detailed information on each SETUP function operation.
3.10.1
<1SELECT>
PURPOSE
To change the active SETUP file number.
Changing the current SETUP file will change the source and in some cases the values of the
variables used by PG7000 in calculating defined pressure. SETUP files should be selected and
changed only by qualified personnel. Incorrect SETUP file selection may cause out of tolerance
calculations and measurements. See Section 3.10, PRINCIPLE, for information on SETUP files and
their use.
OPERATION
To select a SETUP file number to make it the active SETUP file, press
[SETUP]
and select
<1select>
. In the following screen enter the number of the desired SETUP file and press
[ENTER]
. If a valid SETUP file number has been entered, operation returns to the previous
run screen with the selected SETUP file number displayed and active.
Entering an invalid SETUP file number causes an error message to be displayed briefly and
returns to the select SETUP file screen.
If the SETUP file selected specifies COM2 as the source of barometric pressure, PG7000
will attempt to read a barometer on COM2 when initializing the new SETUP file. If PG7000 is
unable to read a barometer on COM2, a communications time out message is displayed briefly and
operation returns to the select SETUP file screen. See Section 3.11.5.4 for information on
configuring COM2 to read an external barometer.
3.10.2
<2VIEW>
PURPOSE
To view the variable source choices and values of any SETUP file number.
OPERATION
Selecting <2view> SETUP allows the variable source choices and values of the active SETUP
file for viewing only. <2view> does not make the viewed SETUP file active. To select the active
SETUP file, press [SETUP] and select <1select>.