See Table 19 for a listing of the source choices for each variable.
To view an existing SETUP file, press
[SETUP]
and select
<2view>
. The display is:
1.
Entry field for number of SETUP file to be viewed. Defaults
to SETUP file currently selected. Must be a number from 2 to
10.
View SETUP file:
#02
Enter the desired SETUP file number and press
[ENTER]
.
If a valid SETUP file number is entered, the display is:
View #02 latmP 2ambT
3%RH 4PCT 5g 6READY
7vac
Select
<1atmP>
to view the atmospheric pressure variable source choice and current value.
The value is displayed in the current pressure units (see Section 3.9.3).
Select
<2ambT>
to view the ambient temperature variable source choice and the
current value. The value is displayed in degrees Centigrade
[
°
C]
.
Select
<3%RH>
to view the ambient relative humidity variable source choice and the current
value. The value is displayed in %RH.
Select
<4PCT>
to view the piston-cylinder temperature variable source choice and the
current value. The value is displayed in degrees Centigrade
[
°
C]
.
Select
<5g>
to view the gravity variable source choice and the current value. The value is
displayed in meters per second squared
[m/s
2
]
.
Select
<6READY>
to view the values of
Ready/Not Ready
status criteria (see Section 3.4).
These include:
<1Pposition>:
Piston position
Ready/Not Ready
limits (see Section 3.4.1). The value is
displayed as a band in millimeters about mid-float position
[
±
mm]
(see Section
3.5).
<2vac> (PG7601 only):
Maximum reference vacuum pressure limit when operating in absolute by
vacuum mode. The value is displayed in Pascal
[Pa]
.
Select
<7vac>
(PG7601 only)
: to view the reference vacuum variable source choice and the
current value. The value is displayed in Pascal
[Pa]
.
3.10.3
<3EDIT>
PURPOSE
To edit an existing SETUP file and/or to create a new SETUP file.
See Section 3.10, PRINCIPLE for information on SETUP files and their purpose.
OPERATION
SETUP file #1 is the factory preferred file and it cannot be edited.
See Table 19 for a listing of the source choices for each variable.