2. INSTALLATION
Page 23
© 1998-2008 DH Instruments, a Fluke Company
•
AMH automated mass handler (optional) placement:
If an AMH is being used, plan electrical and
pneumatic connections to it as well as a location to place it when it is removed from the platform (see
the AMH-38/AMH-100 Operation and Maintenance Manual).
2.3
SETUP
2.3.1
PREPARING FOR OPERATION
Before setting up the PG7000 system, see Section 2.2 for information on site
requirements.
To prepare PG7000 for check out and operation:
Set up the PG7000 Platform (see Section 2.3.1.1).
If an optional AMH-100 mass set is being used, set it up with its mass set (see the AMH-
38/AMH-100 Operation and Maintenance Manual).
Make the system pressure interconnections (see Section 2.3.1.2).
If a manual mass set is being used, set up the manual mass set (see Section 2.3.1.3).
2.3.1.1
SETTING UP THE PLATFORM
To set up the PG7000 Platform proceed as follows:
Place the PG7000 Platform on the site table in the desired orientation.
Though the rear panel is usually in the back, any orientation can be used.
Place the PG7000 Terminal at the desired location.
Connect the PG7000 Terminal to the PG7000 Platform using the 25-pin
cable supplied.
Connect electrical power (85 to 264 VAC, 50/60 Hz) to the PG7000 Terminal
using the power cable supplied. Any grounded power cable with a standard
IEC320-313 connection may be used.
(PG7601 Only)
- Install the vacuum vent valve kit on the vacuum vent port
on the rear of the PG7000 Platform. Refer to the instruction sheet provided
with the vent valve assembly.
Connect the reference vacuum source and shutoff valve to the reference
vacuum port. Take measures to assure that vacuum oil cannot return to the
PG7601. If an optional AMH-38 automated mass handler is being used, the
reference vacuum may be connected to the larger KF40 vacuum connection
on the AMH-38 vacuum chamber.
If an external barometer and/or vacuum gauge is/are being used, establish
communications between the barometer/vacuum gauge and the PG7000
Platform by connecting the external device RS232 port to the PG7000
Platform Com2 port and setting up PG7000 to read and use an external
barometer and/or vacuum gauge (see Section 3.11.5.4, 3.11.5.5). Set the
external barometer head height (see Section 3.11.3.3).
If an automated pressure generation/control component is being used,
establish communications between the automated pressure
generation/control component and the PG7000 Platform by connecting the
generation/control component RS232 port to the PG7000 Platform COM3 port