1. INTRODUCTION
Page 9
© 1998-2008 DH Instruments, a Fluke Company
Piston-cylinder modules designated “tare” are delivered without reported
values and are intended only for use on the “tare” PG7000 in high line differential
pressure measurement mode.
1.2.4.2
PC-7200
For uncertainty in piston-cylinder effective area and typical measurement
uncertainty in pressure defined by the piston gauge, see the piston-cylinder
calibration report and current revision of DHI Technical Note 7920TN01.
PC-7200-100
Sensitivity
1
Reproducibility
2
Typical Drop Rate (50 kg)
2 Pa + 1 ppm
±
5 ppm
0.10 mm/min
PC-7200-200
Sensitivity
1
Reproducibility
2
Typical Drop Rate (50 kg)
4 Pa + 1 ppm
±
5 ppm
0.15 mm/min
PC-7200-500
Sensitivity
1
Reproducibility
2
Typical Drop Rate (50 kg)
10 Pa + 1 ppm
±
5 ppm
0.20 mm/min
PC-7200-1
Sensitivity
1
Reproducibility
2
Typical Drop Rate (50 kg)
20 Pa + 1 ppm
±
5 ppm
0.25 mm/min
PC-7200-2
Sensitivity
1
Reproducibility
2
Typical Drop Rate (50 kg)
40 Pa + 1 ppm
±
5 ppm
0.50 mm/min
1
Sensitivity: The smallest variation in input detectable in output.
2
Reproducibility: Combined long term stability of piston-cylinder effective area and masses.