4. REMOTE OPERATION
Page 171
© 1998-2008 DH Instruments, a Fluke Company
4.5.2
RECOMMENDED SEQUENCE FOR A HOST PROGRAM TO
REMOTELY ENABLE HIGH LINE DIFFERENTIAL MODE
USING THE LAST LINE PRESSURE SETTING
The line pressure must already be established using the steps in the previous example
(recommended sequence for a host program to remotely set a new line pressure and
enable high line differential mode).
Use the “HLDVIEW” command to verify whether the
current line pressure is as expected. Ensure that the reference PG7000 COM2 port is
configured identically to the tare PG7000 COM1 port and that the two COM ports are
connected properly.
Change the measured mode to high line differential:
Send:
“UNIT=UUUUD”
(Request for unit UUUU and high line differential mode)
Reply: “UUUUD” (Unit and measurement mode set complete)
Or
Send:
“MMODE=D”
(Request for high line differential mode)
Reply: “D” (Measurement mode set complete)
High line differential target pressures can now be accomplished using the normal
PG remote commands “PS” and “MS”, or using local operation. The current line
pressure results can be reviewed using the “HLDVIEW” command.