Fig u re 37 - Switch DUT to "ON" for upcoming SF:L2O tests.
Note the purpose of the "STOP BEFORE..." break is to allow you to accommodate power-sensitive
equipment, such as a personal computer or equivalent DUT, which cannot be arbitrarily
disconnected from the AC supply without potentially damaging the DUT. This type of equipment is
much less common than equipment that can be switched "OFF" at any time. The more typical
setting of "STOP BEFORE..." is therefore "NO", in which case the BREAK menus of
Fig u re 36
and
Fig u re 37
will not be displayed.
Many types of medical equipment do not fully disconnect from the AC supply when the power
switch is set to "OFF". Even when "OFF", this type of equipment remains effectively connected to
the AC supply and continues to operate in a standby state, for example, by charging an internal
battery. For this type of equipment, "DUT OFF" leakage tests are required under IEC60601 and
AAMI-ES1 standards.
"DUT OFF" leakage measurements are optionally enabled by the user before the test is performed,
by means of the leakage setup menu. If "DUT OFF" tests are enabled, ES601
Plus
will stop after all
leakage measurements have been completed with DUT power switch in "ON" position. ES601
Plus
will then display instructions as shown in the following figure.
Manual Testing/Chapter 5
#
Page 78
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...