ES601
Plus
OPERATING MANUAL
Fig u re 35 - Break if "STOP AFTER..." is "set to YES".
For IEC60601 and AAMI-ES1 standards, AC supply conditions will be tested in the following
order, according to test conditions initially specified by the user on the leakage setup menu:
1
Reverse polarity; Line 2 (neutral) in single-fault condition (
RP:SF L2O
)
2
Reverse polarity; supply in normal condition (
RP:NC
) or with earth (ground) in
single-fault condition (
RP:SF EO
)
3
Normal polarity; Line 2 in single-fault condition (
NP:SF L2O
)
4
Normal polarity; supply in normal condition (
NP:NC
) or with earth in single-fault
condition (
NP:SF EO
)
At each configuration (1.)-(4.) of the AC supply, measurements of enclosure and earth leakage are
performed. Following these measurements, patient and auxiliary (lead-to-lead) leakage will be
measured if those tests were enabled by the user on the setup menu.
When ES601
Plus
changes the AC supply from SF:L2O to NC, and if "STOP AFTER..." of the
Sequence Controls is "NO" (
Fig u re 31
), ES601
Plus
will pause according to the "POWER-UP
DELAY" setting before proceeding to measure the leakage. This delay of 2 to 9 seconds allows the
DUT to startup up and stabilize before leakage readings are taken. Alternatively, if "STOP
AFTER..." of the Sequence Controls is "YES", ES601
Plus
will change the AC supply from SF:L2O
to NC and wait until you press the
CONTINUE
button:
Manual Testing/Chapter 5
#
Page 76
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...