ES601
Plus
OPERATING MANUAL
AUX1 Terminal (MAP)
:
- CAT I
- Max Voltage:
264 VAC out
- Max Current:
2mA AC out
AUX2 Terminal
:
- CAT I
- Max Voltage:
PE potential
- Max Current:
6.5mA AC, 20mA DC
To DUT Case Terminal
:
- CAT I
- Max Voltage:
PE Potential
- Max Current:
40 A AC
Applied Parts Terminals
:
- CAT I
- Max Voltage:
264 VAC (MAP Test ), 500 VDC (Insulation Test)
- Max Current:
14mA AC @264 VAC, 50mA DC @ 500 VDC
Connect measurement terminals
ONLY
as instructed in this manual, see
Chapter 4, Connecting to ES601
1.5 ECG Simulator
Performance Waveforms
- DC Pulse:
4 seconds
- Square Wave:
2 Hz
- Triangle Wave:
2 Hz
- Havertriangle Pulse:
10, 25, 40, 100 and 200 msec, @ 60 BPM
- Sine Wave:
0.5, 10, 20, 40, 50, 60, 70 and 100 Hz
- Normal Sinus Rhythm: 30, 60, 90, 120, 150, 180, 240 and 300 BPM
Specifications/Chapter 1
#
Page 3
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...