8.5.7 Load Current (IEC60601, AAMI-ES1, VDE0751, IEC62353 and IEC61010)
Load Current is a test where ES601
Plus
connects an internal ammeter to the DUT and measures
the current drawn from the AC line when the DUT is powered and running. This test can be
enabled under the Load Current radio group by selecting Measure, or disabled by selecting Skip.
8.5.8 Protective Earth (IEC60601, AAMI-ES1, VDE0751, IEC62353 and IEC61010)
Protective Earth is a test where ES601
Plus
tries to source current into the dead short circuit that
should exist from the ground pin on the line cord of the DUT to any exposed metal part on the
DUT's enclosure. ES601
Plus
measures the resulting resistance in ohms. The test can be performed
with a 25 amp AC test current (per IEC60601 and IEC61010 Standards), but only if ES601
Plus
contains the optional Protective Earth Test Module (PETM) accessory. Otherwise, the test will
default to a 1 amp test current, or the 1 amp test current can be selected via the radio button. The
Skip option will skip the protective earth test entirely.
8.5.9 Leakage Current (IEC60601 and AAMI-ES1)
If the Leakage Current checkbox is checked, ES601
Plus
will perform Enclosure and Earth Leakage
Current tests. Patient and Auxiliary Leakage Current tests may be optionally enabled. Conditions of
the AC power supply for these leakage tests can be set under the Supply Conditions panel.
8.5.10 Leakage Current (VDE0751 and IEC62353)
If the Leakage Current checkbox is checked, ES601
Plus
will perform a Device Leakage Current test.
A Patient Leakage Current test may be optionally enabled. Note: The AC supply conditions for each
test are: Device Power is On, Line 2 is Normal (connected), and Ground is Normal, according to
the requirements of VDE0751.
8.5.11 Leakage Current (Battery-powered devices)
Under IEC601-BAT, AAMI-BAT, VDE-BAT and IEC353-BAT, which are intended for self-
powered devices, the leakage current test cannot be disabled. The type of test performed is Patient
Leakage Current.
ES601pc Software/Chapter 8
#
Page 148
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...