ES601
Plus
OPERATING MANUAL
Fig u re 38 - Switch DUT to "OFF" for "DUT OFF" tests.
Upon pressing
CONTINUE
on the above menu, ES601
Plus
repeats all leakage tests that were
enabled by the user. Since the DUT power switch remains "OFF" for all measurements, ES601
Plus
does not stop when the AC supply is changed and no further breaks in the sequence occur until the
end of the test.
Once the leakage measurement sequence has stepped through all specified test conditions,
ES601
Plus
will stop and display the last measurement taken during the leakage test. Pressing the
CONTINUE
button on the display saves the leakage measurement data in the Clip Record, along
with the limits corresponding to each measurement. Refer to Chapter 2 for a discussion of the
leakage current protocols of ES601
Plus
, and the data which is saved by each protocol.
Manual Testing/Chapter 5
#
Page 79
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...