6.2.5.6 The R Instruction
The
R
Instruction combines features of the C Instruction (remote control) and the E Instruction
(data entry). The user interface for the R Instruction is the same as that for the E Instruction. The R
Instruction can be useful with patient simulators (e.g. ECG, SpO2, or NIBP) which do not, by
themselves, generate measurement data.
When ES601
Plus
executes the R Instruction, it will transmit the command portion of the
Instruction via the Tester Port specified in the AutoSequence. The prompt portion of the
Instruction will be displayed on the LCD as the "TASK". The ENTER TEST DATA button on the
LCD allows the user to record measurement data or observations via the QWERTY touch keypad,
or by means of a PS/2 keyboard connected to ES601
Plus
.
Automated Testing/Chapter 6
#
Page 123
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...