ES601
Plus
OPERATING MANUAL
Fig u re 60 - ECG AutoSequence in progress.
Throughout the ECG AutoSequence,
PASS
,
FAIL
and
FAULT FOUND AND CORRECTED
buttons are available. "Pass" indicates the DUT reflects the waveform as produced by ES601
Plus
.
"Fail" indicates the waveform is not correctly displayed on the DUT. "Fault Found and Corrected"
indicates the problem causing what would normally be a failure was resolved (such as a bad lead that
was replaced) during the test.
After stepping through the waveform sequence, the QWERTY keypad is displayed to allow
comments to be added to the test, such as “Lead 2 / RL bad – Replaced”. The comments field
allows for 40 ASCII characters. Entry of the comment is optional. When the
ENTER
key on the
QWERTY keypad is pressed, the SAVE TEST RECORD menu is displayed.
Automated Testing/Chapter 6
#
Page 105
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...