ES601
Plus
OPERATING MANUAL
STOP AFTER
SUPPLY ON:
If set to "YES", this control will override the power-up delay setting
described prior and will cause the test to
fu lly s to p
immediately after the
AC supply is connected to the DUT by ES601
Plus
. ES601
Plus
will
prompt the user to activate the DUT and will not proceed until the user
acknowledges by pressing the
CONTINUE
button. This feature
accommodates equipment with a momentary ON/OFF power switch,
which cannot be switched ON until a full AC supply is provided to the
equipment.
STOP BEFORE
SUPPLY OFF:
If set to "YES", this control will cause the test to
fu lly s to p
immediately
before a step which disconnects the AC supply from the DUT.
ES601
Plus
will prompt the user to shut down the DUT and will not
proceed until the user acknowledges by pressing the
CONTINUE
button.
After
CONTINUE
is pressed, ES601
Plus
will then disconnect the AC
supply from the DUT. This feature accommodates equipment having
special shutdown requirements, such as diagnostic imaging equipment or
equipment incorporating a personal computer.
BACK:
Returns to the previous menu (leakage test conditions or applied part
setup, as appropriate).
NEXT:
Displays instructions for connecting the DUT to ES601
Plus
. The
instructions displayed will depend on the safety standard. The example
menu shown following is for the IEC60601 standard.
Manual Testing/Chapter 5
#
Page 72
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...