Fig u re 1 - Startup Menu
3
Setup
3.1 Powering Up/Start-up Menu
When the power to ES601
Plus
is switched on, the Start-up menu is displayed while internal self-tests are
conducted. The Start-up menu displays the firmware version and the current time and date.
Depending on how the "STARTUP MODE" has been configured under System Settings, the above menu
may remain on the LCD indefinitely, or may be automatically replaced by either the MANUAL MODE
menu or the AUTO MODE menu after a few seconds.
Optionally, the "STARTUP MODE" can be set to enable a test of the AC supply voltage at five seconds
after power-on. If the AC supply test at startup has been enabled, ES601
Plus
will produce an alarm if a
reversed-supply condition is detected. Note that a reversed AC supply may affect test results, particularly the
leakage measurements. In the event of a reversed-supply alarm, connect to a different wall outlet or
investigate the wiring of the wall outlet in use before proceeding.
Setup/Chapter 3
#
Page 21
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...