Fig u re 78 - Test Procedure in progress, D instruction.
6.2.5.8 The D Instruction
The D Instruction can be used to provide timed delays in Tester AutoSequences and Test
Procedures (
Fig u re 78
). The delay of the Instruction may range from 1 to 32000 seconds. The
instruction also provides a prompt which is displayed on the LCD as the "TASK" while the timer
counts down.
ES601
Plus
will automatically advance to the next step of the sequence when the timer reaches zero
count. In this case, an "OK" is saved as the test result for the delay step. Alternatively, the user may
press
SKIP
to bypass any remaining delay. The
BACK
button decrements ES601
Plus
sequencer to
the previous step.
Automated Testing/Chapter 6
#
Page 125
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...