Fig u re 32 - DUT connection instructions for IEC60601 test.
According to the CONNECTIONS menu, at minimum an IEC60601 test will require the user to
connect the DUT to the test socket and the Kelvin Cable of ES601
Plus
to the enclosure of the
DUT. The power switch of the DUT must also be set to "ON" before the test is started. Depending
on how the user has chosen to set up the test via preceding menus, additional connections may be
required between AP jacks on ES601
Plus
and DUT applied parts; between AUX1 of ES601
Plus
and DUT signal I/O; and between AUX2 and exposed ungrounded metal on the DUT.
WARNING:
If testing of
T y p e B F
or
T y p e CF
applied parts has been enabled, HIGH VOLTAGE (MAP
voltage) will be periodically output from the AP jacks during the test.
The MAP voltage is isolated from the AC mains and is internally current-limited to 8 milliamperes
maximum at 264 VAC output. Unterminated leads attached to the AP jacks therefore present a
minor, non-lethal shock hazard to the careless user.
Always take care to make proper electrical connections to the DUT before starting a leakage test.
Disconnect all unused leads or cables from the AP jacks on ES601
Plus
.
Manual Testing/Chapter 5
#
Page 73
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...