ES601
Plus
OPERATING MANUAL
ISOLATED
SIGNAL I/O, cont'd:
Therefore, the "ISOLATED SIGNAL I/O" button enables a
supplementary test condition for touch current, patient and auxiliary
leakage measurements under IEC60601 and AAMI safety standards.
The expected result is these leakages should not be affected by
applying mains voltage to the data connection, or by earthing the data
connection, since the input/output signals are supposedly isolated
from the DUT enclosure and patient leads.
WARNING:
If ISOLATED SIGNAL I/O is set to "TEST!", a warning message is displayed: (!) CHECK
D.U.T. SPECS.
When enabling this test, users must be sure the data interfaces being tested are
fu lly is o late d
.
Otherwise,
s e v e re d am ag e to th e DUT
could result when AUX1 is activated during the test. If
in doubt, contact the DUT manufacturer or supplier for guidance.
Datrend Systems Inc. will not be responsible for damage resulting from incorrect application of
this test.
UNGROUNDED
METAL:
Options are "DO NOT TEST" (default) or "TEST". This parameter
determines whether the AUX2 output of ES601
Plus
will be connected
to protective earth of the AC supply.
AUX2 is intended to connect to
flo atin g e xp o s e d
metal on the DUT.
This test is required under the IEC60601 standard if the DUT has
metal parts that are electrically isolated or "floating", but are exposed
so a user or patient can touch them. If enabled, the test does
NOT
measure the leakage current from the floating metal parts. In this test,
the metal is connected to protective earth and then the leakage current
of the
patient-applied parts
is measured.
Therefore, this button enables a supplementary test condition for
patient leakage tests per IEC60601. The expected result is the patient
leakage is unaffected by connecting such metal parts to earth, since the
metal is supposed to be electrically isolated.
Manual Testing/Chapter 5
#
Page 66
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...