Fig u re 31 - Sequence Controls for manual leakage test.
The buttons provided on the Sequence Control menu are:
PAUSE AFTER
MEASUREMENT:
After each measurement of the test, ES601
Plus
will pause for the
indicated time before proceeding with the next leakage measurement. The
delay provided by this control allows the user to more easily interrupt a
sequence in progress, so a particular leakage measurement of interest can
be repeated as desired. This control can be set to "NO" to disable the
pause.
POWER UP
DELAY:
During the test, ES601
Plus
will wait for the indicated time when the AC
supply is applied to the DUT, before proceeding with the sequence. This
delay allows the DUT to start up and stabilize before subsequent leakage
measurements are taken. The available options are 2 to 99 seconds in one
second steps.
Manual Testing/Chapter 5
#
Page 71
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...