ES601
Plus
OPERATING MANUAL
7.4.2 Safety Test Execution
Under computer control, a Safety AutoSequence is initiated by means of this exchange:
a.
PC software transmits
RSAF\r
to ES601
Plus
via the serial connection.
b.
ES601Plus responds by returning
*\r
(0x2A 0x0D) to the PC. If the
RSAF\r
command is received by ES601
Plus
without
errors, the
*\r
response should be returned to the PC within 0.2 seconds.
c.
PC software sends one ASCII character to ES601
Plus
to specify the AutoSequence.
Send:
a
(0x61) to select AutoSequence #1; or
b
(0x62) to select AutoSequence #2; or
c
(0x63) to select AutoSequence #3;
...
and so on up to
x
(0x78) to select AutoSequence #24. PC software must send this
character within two seconds of the
*\r
response from ES601
Plus
, otherwise
ES601
Plus
will time out and abort the protocol. If ES601
Plus
receives a character
other than
a
to
x
, ES601
Plus
returns
?\r
to the PC and aborts the protocol.
As an alternative to sending a lower-case letter, PC software can alternatively send an
upper case letter:
A
(0x41) specifies AutoSequence #1,
B
(0x42) specifies
AutoSequence #2, and so on up to
X
(0x58) which specifies AutoSequence #24. The
difference is that sending lower-case causes ES601
Plus
to execute a complete
AutoSequence as if the test had been selected from the SAFETY
AUTOSEQUENCE menu, beginning with CONNECTION instructions,
APPLIED PARTS setup, and SEQUENCE control menus. Alternatively, for
upper-case letters, these initial information and setup menus are bypassed and
execution of the test begins immediately with the first measurement.
Computer Control/Chapter 7
#
Page 135
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...