6.2.3 Tester AutoSequences
ES601
Plus
provides automated test sequences for medical device analyzers and testers such as IV
Pump Testers, Pulse Oximeter Testers, Defibrillator Testers, ESU Testers, NIBP Testers, plus many
others. To accommodate any make or model of test equipment you may own, ES601
Plus
provides
five sets of User-Defined Tester AutoSequences, which you can arbitrarily program as desired using
the AutoSequence Editor of ES601pc. For each of the User-Defined Testers, ES601
Plus
provides
you with 12 programmable AutoSequences.
A Tester AutoSequence can comprise from 1 to 24 steps, where each step corresponds to an
ES601
Plus
Instruction. Typically, Tester AutoSequences will be composed of ES601
Plus
re m o te
c o n tro l
Instructions. In setting up a Tester AutoSequence, you will also select which ES601
Plus
T e s te r Po rt
will interface with your external analyzer, and the
b au d rate
to be used for RS232 serial
communications with your analyzer. These selections are made with the AutoSequence Editor of
ES601pc.
A Tester AutoSequence cannot include an ES601
Plus
Safety AutoSequence or ECG AutoSequence;
to include a safety and/or ECG test within the scope of another automated test requires you to set
up a full Test Procedure.
6.2.4 Test Procedures
Test Procedures are created using the Test Procedure Editor of the ES601pc software package. A
Test Procedure can comprise from 1 to 96 steps, where each step corresponds to an ES601
Plus
Instruction.
In addition to individual Instructions, Test Procedures can also incorporate General Inspections and
Tester AutoSequences, provided the total number of steps created by including such automated tests
in a Procedure does not exceed 96. Additionally, any
o n e
of the Test Procedure steps can optionally
be a Safety AutoSequence, and any
o n e
of the steps can optionally be a ECG AutoSequence. A Test
Procedure cannot have more than one Safety AutoSequence, or more than one ECG AutoSequence.
Automated Testing/Chapter 6
#
Page 117
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...