ES601
Plus
OPERATING MANUAL
Fig u re 33 - Leakage current test in progress.
Once the leakage test has been set up and the connections are made, the test is started by pressing
the
CONTINUE
button on the CONNECTIONS menu. The required measurements are
performed automatically in a sequence. AC RMS and DC currents are measured simultaneously, and
the display is updated with new data approximately two times every second. If the "PAUSE
AFTER..." Sequence Control is not set to "NO", ES601
Plus
will delay an additional 1 to 5 seconds
between consecutive leakage measurements.
The type of measurement being performed and the AC supply conditions are shown on the display,
above the leakage readings. RMS and DC limits corresponding to the test conditions are also shown
on the right, above the readings. Each RMS and DC value is respectively compared to its limit and is
either "PASSED" or "FAILED" automatically by ES601
Plus
. The
PAUSE
button may be pressed
at any time to pause the sequence and freeze the most recent measurement on the display (see
Fig u re 34
).
Manual Testing/Chapter 5
#
Page 74
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...