ES601
Plus
OPERATING MANUAL
Fig u re 36 - Break if "STOP BEFORE..." is set to "YES".
The leakage sequence proceeds with NC tests after you press
CONTINUE
. Alternatively, you may
press
ESCAPE
to abort the leakage test and return to the MANUAL MODE menu. Note the
purpose of the "STOP AFTER..." break is to allow you activate a DUT which cannot be switched
ON until a full AC supply is provided to the DUT.
When ES601
Plus
changes the AC supply from NC to SF:L2O, ES601
Plus
will not stop if "STOP
BEFORE..." of the Sequence Controls is "NO" (
Fig u re 31
). If "STOP BEFORE..." is "YES",
ES601
Plus
will stop and prompt you to shut down the equipment before supply Line 2 is
disconnected:
After
CONTINUE
is pressed, ES601
Plus
will disconnect Line 2 from the DUT and will then
prompt you to reset the DUT power switch to "ON". This is necessary to satisfy the requirements
of the leakage test, but the instruction can be ignored if the DUT is of the type that cannot be set to
"ON" unless a full AC supply is provided.
Manual Testing/Chapter 5
#
Page 77
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...