ES601
Plus
OPERATING MANUAL
Fig u re 23 - Insulation test result.
WARNING:
If testing insulation resistance of applied parts, HIGH VOLTAGE (500 VDC) will be output from the
AP jacks for a few seconds during the test.
The high voltage is internally current-limited to 2 milliamperes maximum. Unterminated leads attached to
the AP jacks therefore present a minor, non-lethal shock hazard to the careless user.
Always take care to make proper electrical connections to the DUT before starting an insulation
resistance test. Disconnect all unused leads or cables from the AP jacks on ES601
Plus
.
After approximately two seconds, the measurement result in megohms is displayed.
On the insulation test result screen, select one of these three options:
Manual Testing/Chapter 5
#
Page 53
Summary of Contents for ES601 Plus
Page 1: ...Operating Manual ES601Plus Inte rnatio nal Auto m ate d Ele c tric al Safe ty Analyze r ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 19: ...ES601Plus OPERATING MANUAL Definitions Page xv ...
Page 39: ...ES601Plus OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 57: ...ES601Plus OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 117: ...ES601Plus OPERATING MANUAL Manual Testing Chapter 5 Page 98 ...
Page 151: ...ES601Plus OPERATING MANUAL Automated Testing Chapter 6 Page 132 ...
Page 159: ...ES601Plus OPERATING MANUAL Computer Control Chapter 7 Page 140 ...
Page 193: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 174 ...
Page 195: ...ES601Plus OPERATING MANUAL ES601pc Software Chapter 8 Page 176 ...
Page 223: ...ES601Plus OPERATING MANUAL Appendix B Page 204 ...
Page 227: ...ES601Plus OPERATING MANUAL Appendix C Page 208 ...
Page 228: ...ES601Plus OPERATING MANUAL APPENDIX D Importing Equipment Records Appendix D Page 209 ...
Page 229: ...ES601Plus OPERATING MANUAL Appendix D Page 210 ...