Appendix F
In-Sight 1700 Wafer Reader Mount Conversion
The In-Sight 1700 wafer reader can be mounted either horizontally or vertically
depending on the positions of the read window and light trap (see Figure F-1). The
In-Sight 1700 wafer reader can be easily converted from one mounting configuration to
the other. In the following example, a horizontal mount In-Sight 1700 wafer reader is
converted to a vertical mount.
NOTE
If possible, perform these steps within a cleanroom environment (Class 1000 or better) at an
electrostatic discharge (ESD)-safe workstation. Use cleanroom-approved gloves to avoid
fingerprints on the read window. If the conversion is not done in a cleanroom, there is a possibility
of introducing "artifacts" on the optics. Also, depending on the Class of the cleanroom where the
wafer reader will be used, the wafer reader may no longer meet Class specifications after the
conversion has been done.
To convert the mount configuration
:
1
Using a .050 inch Allen wrench, remove the two 2-56 x 3/16" button head screws that
attach the light trap, and the two screws that attach the read window.
Figure F-1: Location of Light Trap and Read Window (Horizontal Mount Configuration)
141
Summary of Contents for In-Sight 1721
Page 1: ...IN SIGHT 1700 SERIES WAFER READER USER MANUAL...
Page 2: ......
Page 8: ...In Sight 1700 Series Wafer Reader vi...
Page 58: ...Section 4 Managing Job Files In Sight 1700 Series Wafer Reader 48...
Page 130: ...Section 8 Specifications In Sight 1700 Series Wafer Reader 120...
Page 136: ...Updating the Wafer Reader s Firmware In Sight 1700 Series Wafer Reader 126...
Page 140: ...Configuring Microsoft Windows Network Settings In Sight 1700 Series Wafer Reader 130...
Page 154: ...In Sight 1700 Wafer Reader Mount Conversion In Sight 1700 Series Wafer Reader 144...
Page 156: ...In Sight 1700 Series Wafer Reader www cognex com products InSight default asp Rev 16 May 2006...