In-Sight
®
1700 Series Wafer Reader
Updating the Wafer Reader’s Firmware
5
Click Login to log on to the In-Sight sensor's FTP file system. If the logon is
successful, all of the files currently installed on the wafer reader are displayed in the
Internet Explorer window, as shown in Figure A-2.
Figure A-2: Installed Wafer Reader Files
6
Open a separate Microsoft Windows Explorer window and locate the .HTM, .JAR,
WafID00.JOB, and .SET files in the In-Sight PC Host installation directory. The
default directory is: “C:\Program Files\Cognex\In-Sight\In-Sight Wafer ID (Java)
3.x.x”.
7
Copy and paste (or drag-and-drop) the files located in the In-Sight PC Host
installation directory to the wafer reader’s FTP file system, refresh the view; the
selected files have been copied to wafer reader flash memory and now appear in the
browser window, as shown in Figure A-3.
Figure A-3: Installed Wafer Reader Files