10. Before Calling a Service Engineer
– 113 –
10.2 Alert
Error
Code
Error Message (Upper box: Error type
Lower box: Error description)
Remedy
-4
System Info Error
Touch
[OK]
to close the window, and then see
(
→
,
and turn off the CXDI.
When these errors occur, note the error code
and contact a Canon representative or
distributor.
Driver info is updated.
If there is the trouble of sensor, the sensor power unit may not be supplied.
-5
System Info Error
Capture device info is updated.
If there is the trouble of sensor, the sensor power unit may not be supplied.
-6
System Info Error
A/D board info is updated.
If there is the trouble of sensor, the sensor power unit may not be supplied.
-7
System Info Error
LANMIT info is updated.
If there is the trouble of sensor, the sensor power unit may not be supplied.
-8
System Self Check Error: (Contents)
Call a service engineer and let him/her know the error code.
-115
Capture Board Error
Detected Capture Board Error.
-116
Sensor Unit: Self Check Error
Detected an Error.
-119
Image Error
Touch
[OK]
to close the window.
Perform operations according to the messages
displayed.
Note the error code and contact a Canon
representative or distributor.
An error occurred in the image. Call a service person. Check the image in the
Preview-screen and also in the high-resolution monitor or by using [MAG.] or
[QA-MAG.] in the QA screen. Also, turn the power box off and on again to
restart the CXDI.
-546
Internal Storage Limit Warning:
Touch
[OK]
to close the window.
The amount of available internal storage space has dropped below (current
settings). Images currently in storage commitment or that failed in storage
commitment will be automatically deleted.
-547
Internal Storage Limit Warning:
Touch
[OK]
to close the window and then see
, and
delete the image.
The amount of available internal storage space has dropped below (current
settings). Check the contents of the study list.
-5002
OPU-CCR RPC Error
Touch
[OK]
to close the window, and then see
(
→
,
and turn off the CXDI.
When these errors occur, note the error code
and contact a Canon representative or
distributor.
OPU-CCR RPC Error.
-6001
OPU-CCR RPC Error
Operation has been canceled.
Summary of Contents for CXDI-1
Page 9: ...Starting Up Shutting Down 1 Starting Up Shutting Down the CXDI ...
Page 15: ...Exposure and Output 2 Taking an Image 3 Output Control Process Viewer ...
Page 38: ...2 Taking an Image 30 ...
Page 60: ...5 Viewing Multiple Images Multi View Screen 52 ...
Page 70: ...6 Browsing Completed Studies Study List 62 ...
Page 71: ...Processing the Image 7 Processing the Image QA Mode Screen ...
Page 103: ...CXDI Adjustment 8 Calibration 9 Self test ...
Page 110: ...8 Calibration 102 ...
Page 116: ...9 Self test 108 ...
Page 117: ...Troubleshooting 10 Before Calling a Service Engineer 11 Questions and Answers ...
Page 136: ...11 Questions and Answers 128 ...
Page 156: ...B Using the Rejection Reason Input Function 148 ...
Page 162: ...C Using the Sensor Unit Attach Detach Function 154 ...
Page 169: ......