8. Calibration
– 100 –
3
Exposure is completed.
When the last exposure is completed, the message
“Please wait.”
is displayed, and data
processing is started.
When data processing results in an error
A window is opened, and an error message is displayed.
After checking the error details, touch
[OK]
to close the window.
Touch [
EXIT]
to cancel calibration, or
[RETAKE]
to perform calibration again.
When the grid stripes have been detected in the course of the calibration of
the CXDI-40G COMPACT
A window is opened, and an error message is displayed.
After checking the error details, touch
[OK]
to close the window.
To cancel calibration, touch
[EXIT]
; to perform calibration again, check that the grid
has been removed, and then touch
[RETAKE]
.