– 149 –
C. Using the Sensor Unit Attach/Detach Function
NOTE:
This function can be used only with sensor units equipped with an attach/
detach function.
Be sure to also refer to the user’s manuals for both the sensor unit and power
box on the use of this function.
The procedure below is an example of operations when using the category type exposure screen.
C.1 Attach/Detach Function of the Sensor Unit
The sensor unit can be attached/detached while the system is running.
The procedure for attaching and detaching the sensor unit is simple.
NOTE:
To attach/detach the sensor unit, always turn off the power of the power box
first.
NOTE:
Do not remove the connector or disconnect the power of the power box while
image data is being transferred. Image data that has been taken will be lost.
NOTE:
Always attach/detach the sensor unit from the exposure screen.
Attaching/detaching the sensor unit while on another screen may result in
malfunction or loss of data.
NOTE:
If a dialog box is being displayed on the exposure screen, close the dialog
box first and then perform attachment/detachment.
NOTE:
If the sensor unit is detached during READY status, READY status will be
canceled, and the selected exposure mode and parameters set on the
parameter changing screen will become disabled. To continue taking
exposures, select the exposure mode and make settings again (
after connecting the detached sensor unit.
Summary of Contents for CXDI-1
Page 9: ...Starting Up Shutting Down 1 Starting Up Shutting Down the CXDI ...
Page 15: ...Exposure and Output 2 Taking an Image 3 Output Control Process Viewer ...
Page 38: ...2 Taking an Image 30 ...
Page 60: ...5 Viewing Multiple Images Multi View Screen 52 ...
Page 70: ...6 Browsing Completed Studies Study List 62 ...
Page 71: ...Processing the Image 7 Processing the Image QA Mode Screen ...
Page 103: ...CXDI Adjustment 8 Calibration 9 Self test ...
Page 110: ...8 Calibration 102 ...
Page 116: ...9 Self test 108 ...
Page 117: ...Troubleshooting 10 Before Calling a Service Engineer 11 Questions and Answers ...
Page 136: ...11 Questions and Answers 128 ...
Page 156: ...B Using the Rejection Reason Input Function 148 ...
Page 162: ...C Using the Sensor Unit Attach Detach Function 154 ...
Page 169: ......