9. Self test
– 104 –
4
Start the self test.
Touch
[TEST START]
.
The exposure parameters for the last calibration are displayed. Set the exposure
parameters of the x-ray generator to match the displayed parameters.
Ready State
5
Press the exposure button.
Press the exposure button on the x-ray generator. The self test is started.
The self test consists of five parts from
TEST1 to TEST5
. The self test is finished when
TEST5
is completed.
When the test is successfully completed,
Passed.
is displayed for
TEST1 to TEST5
.