7. Processing the Image (QA Mode Screen)
– 90 –
7.9 CORRECT PROC.
In this screen, the dosage-dependent sharpness of the image can be set, the stripes caused by the
fixed grid can be decreased, and the collimation mask can be set.
CORRECT PROC. Screen
7.9.1 Setting the Dosage-Dependent Sharpness Adjustment
1
Enable the dosage-dependent sharpness.
Touch and highlight (white character)
[D. DEPEND MTF. IMP]
.
2
Set the type.
Touch to open the window and select the body part.
Type Button Window
3
Set the intensity as required.
Touch or to select a setting from
0 to 3
.
The displayed value on the left is the saved value, and the value on the right is the modified
value.
Larger values provide more sharpness.