85
MD 610_2d 11/2019
10. Closethe vial tightly with the cap and swirl severaltimes
until the tablets are dissolved.
11. Placethe vial in the sample chamber making sure that
the
marks are aligned.
12. Press
TEST
key.
Record the displayed test result (G).
13. Removethe vial from the sample chamber, empty the
vial, rinse vial and cap several times. Fill with
a few
drops of water sample
.
14. Add
one DPD No. 1 tablet
straight from the foil and
crush the tablet using a clean stirring rod.
15. Add water sample to the 10 ml mark.
16. Closethe vial tightly with the cap and swirl severaltimes
until the tablet is dissolved.
17. Placethe vial in the sample chamber making sure that
the
marks are aligned.
18. Press
TEST
key.
Record the displayed test result (A).
19. Removethe vial from the sample chamber.
20. Add
one DPDNo. 3 tablet
straight from the foil to the
same water sample and crush the tablet using a clean
stirring rod.
21. Closethe vial tightly with the cap and swirl severaltimes
until the tablet is dissolved.
22. Placethe vial in the sample chamber making sure that
the
marks are aligned.
Zero accepted
prepare Test
press TEST
Zero accepted
prepare Test
press TEST