82
MD 610_2d 11/2019
Chlorine HR (Kl)
with Tablet
5 – 200 mg/l Cl
2
Insert the adapter for 16 mm Ø vials.
1.
Fill a clean vial (16 mm Ø) with
8 ml of the water
sample
, closetightly with the cap.
2.
Placethe vial in the sample chamber making sure that
the marks are
l
aligned.
3.
Press
ZERO
key.
4.
Removethe vial from the sample chamber.
5. Add
one CHLORINE HR (KI) tablet
straight from
the foil to the water sample and crush the tablet using
a clean stirring rod.
6. Add
one ACIDIFYING GP tablet
straight from the
foil to the samewater sampleand crushthe tablet using a
clean stirring rod.
7.
Closethe vial tightly with the cap and swirl severaltimes
until the tablets are dissolved.
8.
Place the vial in the sample chamber making sure
that the marks are
l
aligned.
9.
Press
TEST
key.
The result is shown in the display in mg/l Chlorine.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Ø 16 mm
0
5
1